| Title | SKU | Price |
|---|---|---|
| Default Title | AI:H200-15sst.220 |
$2,590
|
The Ai H200 stainless steel series heated recirculators with 2,800 watts heater offer outstanding performance and high reliability as well as quiet operation and ease-of-use. Powerful force/suction pump provides constant flow rates, while the 2nd generation PID controller gives you the ideal solution when you need precise temperature control.
Features:
| Type | Stirred / Circulating Baths |
| Incubation | Yes |
| Refrigeration | No |
| Temp Range (°C) | Ambient to 200°C |
| Capacity (L) | 15 L |
| Circulation | External Only |
| Brand | Across International |
| MaxFlow Rate (l/min) | 15 |
| Temp Stability (°C) | ±2°C |
| Timer | Yes |
| Alarms | Yes |
| Digital or Mechanical | Digital |
| Data connection | No |
| Tank Material | Stainless Steel |
| Sample Agitation | No |
| Media Racks Available | No |
| Warranty (years) | 1 |
| Weight (lb) | 40 |
| Width (in) | 10 |
| Depth (in) | 12 |
| Height (in) | 8 |
| Model | H200-15 |
| Temperature range | Ambient to 200°C |
| Electrical requirements | 220V 60Hz 1-Phase 2,900 watts |
| Heating capacity | 2,800 watts |
| Recirculating pump | Pumping speed: 15 liters per minute, adjustable Maximum head: 16 feet or 5 meters Power: 100 watts |
| Reservoir | Dimensions: 10 x 12 x 8" (WxDxH) Material: Type 304 stainless steel Capacity: 15 liters or 4 gallons Thermo fluid connection: M16 x 2 Fluid drain port: 3/8" hose barb with on/off valve |
| Temperature controller | Type: Digital microprocessor PID controlled Display: 7" touch screen Display unit: Celsius °C Temperature step: 0.1°C Temperature accuracy: +/- 2°C Temperature range: Ambient to 200°C |
| Working environment | -15°C/5°F to 25°C/77°F, at <66% humidity |
| Unit/shipping dimensions | 14.75 x 19 x 18.25" / 17 x 22 x 22" (WxDxH) |
| Unit/shipping weight | 40 Lbs / 55 Lbs |
| Safety | Short circuit and ground protection Over-current protection Over-temperature protection Upper/lower temperature alarm protection |
| Warranty | One year |
| Compliance | UL 61010-1 :2012 Ed.3+R:29Apr2016 CSA C22.2#61010-1 :2012 Ed.3 |