| Title | SKU | Price |
|---|---|---|
| Default Title | AI:M14 |
$2,840
|
Five point temperature sensor system, UL certified, built-in USB data port. Great for vaccine and medicine storage
Features:
Please note: This unit is NOT designed to be used on a mobile platform. Unit is preset to 5°C as per CDC guidelines for vaccine storage.
| Min. Temperature | 2°C to 8°C |
| Cubic Feet | 13.8 |
| Alarms | Yes |
| Controls | Digital |
| Defrost | Manual |
| Electrical | 110V (±10%) 60Hz |
| Warranty(years) | 1-year |
| Weight (lb) | 256 |
| Width (in) | 24.5 |
| Depth (in) | 26 |
| Height (in) | 78.8 |
| Model | M14 |
| Electrical requirements | 115V (±10%) 60Hz 1-Phase 240W Door heater: 90W LED: 3W |
| Temperature range | 2°C to 8°C, with ±2°C uniformity |
| Compressor |
Number of compressor: one Make: Embraco Refrigerant: R600a |
| Chamber |
Dimensions: 22 x 18 x 53" (WxDxH) Capacity: 13.8 Cu Ft Lighting: 3 watts low heat emission LED |
| Shelf | Material: powder coated steel Quantity: 7 |
| Door | Glass, heated (90W heater) |
| Controller | USB data export Backup battery: 48 hours |
| Cooling | Forced air circulation |
| Working environment | - Temperature: 16°C to 32°C (ideal temperature: 18°C to 25°C) - Humidity: <85% Rh - Altitude: < 2000 m / 6561 Ft - Stable, well-ventilated platform with no excessive dust - Keep freezer at least 12 inches from any walls or ceiling - Use dedicated power, do not share outlet - Extension power cable should be less than 10 feet is required - Freezer and power outlet should be grounded properly - Avoid direct sunlight, water spill and do NOT use freezer outdoor - Clean condenser fins and remove frost inside chamber at least once a month - Important! Do not plug in your unit. Leave your Ai refrigerator or freezer for 24 hours before turning the unit on for the first time. |
| Access port | One, 1" diameter, top right side of refrigerator |
| Casters | 4 + 2 leveling feet |
| Unit weight & dimensions | 256 Lbs 24.5 x 26 x 78.8" (WxDxH) |
| Safety | Power failure protection Sensor failure protection High/low temperature alarms Condenser failure alarm Backup battery low alarm Door open alarm |
| Warranty | One year |
| Compliance | UL 61010-1 CSA C22.2 |